SNA-26M Spectrograph and amplitude frequency analyzer
|
price: 17000 nominal dollars
|
|
* Spectrograph in the range of 9 kHz - 26,5GHz * Frequency resolution 1Hz * YIG preselector * IF filter with FFT from 10Hz * Scalar analyzer in the range of 5MHz - 26,5GHz * Synthesized microwave oscillator * Output attenuator 0 - 55 dB * Reference source frequency instability ±2*10-8 * Short frequency setting time
|
Display
|
TFT COLOR LCD
|
Diagonal size
|
17 cm (6.5")
|
YIG indicator resolution
|
640(H)x480(V)
|
Quantity of colours
|
16 bit
|
Input/output connector
|
N(m), 50 Ω
|
Supply voltage (alternating)
|
200 - 250 V
|
Mains power consumption
|
≤ 50 W
|
Dimensions
|
320x145x310 mm
|
Weight
|
8 kg
|
Ambient temperature
|
0...40о С
|
Measured frequency range:
|
9kHz - 26.5GHz
|
Subranges
|
Band 0:
|
9 kHz - 2.15 GHz
|
Band 1:
|
2GHz - 26.5 GHz
|
YIG preselector range
|
2 - 26.5 GHz
|
Frequency resolution
|
1 Hz ...
|
IF band pass (on a -3dB level)
|
10Hz - 1MHz
|
IF band error
|
± 10%
|
IF filter selectivity
|
≤ 5:1 (-60dB : -3dB)
|
Videofilter
|
1Hz - 1MHz (1-3 step)
|
Swath
|
0Hz, 1kHz...26.5GHz
|
Reference source frequency
|
60 MHz
|
Frequency instability from temperature
|
± 2*10-8 (0...50oC)
|
Measured level range
|
noise +20 dBm
|
Maximum input level (RF ATT=30dB, ±0Vac)
|
+20 dBm
|
Average level of displayed noises (IF filter 10Hz)
|
-135dBm + 2f (GHz)
|
Input attenuator (step 1dB)
|
Band 0:
|
0...-60 dB (step 1dB)
|
Band 1:
|
0...-55 dB (step 5dB)
|
Attenuator error
|
Band 0:
|
± 0,3dB+5% from RF ATT
|
Band 1:
|
± 2,5dB
|
Level measurement error
|
± 3dB
|
Second harmonic suppression
|
-60 dBc
|
Selectivity over image channel
|
-70 dBc
|
Intermodulation distortions of the 3d degree (two signals with a -27dBm level diversed within 100kHz)
|
-70dBc (level on a mixer -30dBm), distance between two signals >50kHz
|
Scaling point 1dB
|
-10 dBm (RF ATT=0dB)
|
Scale of indicator face (10 points)
|
10, 5, 2, 1 dB/point
|
Unit measure
|
dBm, dBW, dbV, dBmV, dBµV
|
Detection mode
|
Normal, Posi-peak, Nega-peak, Sample, Average (RMS, Video)
|
Frequency range
|
5MHz - 26,5GHz
|
Subranges
|
Band 0:
|
5MHz - 8GHz
|
Band 1:
|
8GHz - 20GHz
|
Band 2:
|
20GHz - 26.5GHz
|
Frequency resolution
|
1 Hz ...
|
Swath
|
from 1kHz to 26,5GHz
|
Measuring limits
|
Band 0:
|
50 dB
|
Band 1:
|
30 dB
|
Band 2:
|
30 dB
|
Output attenuator
|
Band 0:
|
0...-60 dB (step 1dB)
|
Band 1:
|
0...-55 dB (step 5dB)
|
Band 2:
|
0...-55 dB (step 5dB)
|
Attenuator error
|
Band 0:
|
±3dB + 5% from RF ATT
|
Band 1:
|
± 2.5 dB
|
Band 2:
|
± 2.5 dB
|
Measuring error
|
≤± 2dB
|
Switchable scale of indicator scale
|
Band 0:
|
5, 3, 2, 1 dB/point
|
Band 1:
|
3, 2, 1 dB/point
|
Band 2:
|
3, 2, 1 dB/point
|
Reference signal level
|
0dBm (OUT ATT = 0dB)
|
Maximum sweep time
|
Band 0:
|
150 ms
|
Band 1:
|
300 ms
|
Band 2:
|
300 ms
|
Frequency range (with application of external directional couplers)
|
5MHz - 26.5 GHz
|
Subranges
|
Band 0:
|
5MHz - 8 GHz
|
Band 1:
|
8GHz - 20 GHz
|
Band 2:
|
20GHz - 26,5 GHz
|
Swath
|
from 1 kHz to 26.5GHz
|
Frequency resolution
|
1 Hz ...
|
Measuring limits
|
Band 0:
|
1.07 - 5
|
Band 1:
|
1.2 - 5
|
Band 2:
|
1.2 - 5
|
Measuring error not worse than
|
± (4K+5)%, where K is a measured value
|
Scale of indicator face
|
Band 0:
|
1.07 - 5 / 3.6 / 1.9
|
Band 1:
|
1.2 - 5 / 3.6 / 1.9
|
Band 2:
|
1.2 - 5 / 3.6 / 1.9
|
Sweep-time
|
Band 0:
|
150 ms
|
Band 1:
|
300 ms
|
Band 2:
|
300 ms
|
Frequency range
|
5MHz - 26.5 GHz
|
Subranges
|
Band 0:
|
5MHz - 8GHz
|
Band 1:
|
8GHz - 20GHz
|
Band 2:
|
20GHz - 26.5GHz
|
Frequency resolution
|
1 Hz ...
|
Phase noise (while tuning out by 100kHz)
|
≤ -95 dBc/Hz
|
Reference source frequency
|
60 MHz
|
Frequency instability from temperature
|
±2*10-8 (0...50oC)
|
Maximum output level
|
0dBm (OUT ATT = 0dB)
|
Output attenuator
|
Band 0:
|
from 0 to -60 dB (step 1dB)
|
Band 1:
|
from 0 to -55 dB (step 5dB)
|
Band 2:
|
from 0 to -55 dB (step 5dB)
|
Attenuator error
|
Band 0:
|
±3dB + 5% from RF ATT
|
Band 1:
|
±2.5dB
|
Band 2:
|
±2.5dB
|
Output level setting error not worse than
|
± 4dB
|
Second harmonic suppression not worse than
|
Band 0:
|
-50 dB
|
Band 1:
|
-20 dB
|
Band 2:
|
-20 dB
|
Compound signal suppression not worse than
|
-50 dBc
|
Maximum frequency setting time
|